Volume 61, Number 1, 63-66, DOI: 10.1134/S1061934806010138

High-performance liquid chromatographic determination of simvastatin in medical drugs

Eyad S. M. Abu-Nameh, Reyad A. Shawabkeh and Azzam Ali

View Related Documents

Abstract

A simple and rapid HPLC method for the determination of simvastatin using a C18-Hypersil column and acetonitril-phosphate buffer-methanol (5: 3: 1, v/v/v) as a mobile phase with detection at 230 nm was proposed. Commercial pharmaceutical tablets were analyzed with a linear range for simvastatin up to 1.884 mg % and a regression coefficient of 0.9995. The method is found to be precise, accurate, reliable, and selective.
The text was submitted by the authors in English.