Purpose
Corneas harvested post-mortem are at risk of contamination, therefore antibiotic additives are used in cold storage and organ
culture systems. In the latter, sterility testing of the medium is part of the standard protocol. Intuitively, testing after
longer organ culture periods should be more likely to detect contaminations than early testing, but may delay allocation.
This study evaluates whether an optimal time for detection of donor cornea contamination can be identified.
Methods
The study complies with the Declaration of Helsinki. All procedures were supervised using a certified quality management system
(ISO 9001:2000). Donor corneas harvested by enucleation or corneoscleral excision over 5 consecutive years were processed
according to German and EC laws and guidelines. The corneas were stored in a closed organ culture system in 100 ml MEM containing
penicillin, streptomycin and amphotericin B at 31°C for up to 28 days without media exchange. In 762 corneas, 10 ml samples
of medium, obtained between days 3 and 8 of culture, were tested for sterility in an automated detection system (BacT/ALERT,
bioMérieux). In 424 corneas, a second sterility test was performed from the same medium before release. Contamination detection
probabilities were related to the culture duration before the primary test (Cochran–Armitage).
Results
Overall, 19 contaminations were found. Contaminations were bilateral in four donors. One contamination was apparent by macroscopic
inspection of the medium prior to the primary sterility test; 12 were detected upon primary sterility testing. Furthermore,
six primarily undetected contaminations were observed: five in the secondary sterility test and one suspected microscopically.
In most cases, contamination could also be seen by medium turbidity and acidification, but in three cases macroscopic medium
changes were significantly delayed or absent. No trends were found between the times of primary sterility sampling and both
positive and false negative test outcome probabilities.
Conclusion
Detection probability of contaminations in organ culture media does not increase between days 3 and 8; therefore, sterility
can be tested on day 3. With the recommended follow-up after sterility testing being 7 days, microbiologic release can take
place after 10 days of culture. Nevertheless, the testing is not failsafe, and should always be combined with macroscopic
inspection of the media.
Keywords Cornea - Organ culture - Transplant - Microbiologic testing