Introduction
Current ambulatory ECG monitoring systems are limited in their ability to diagnose patients with palpitations. The aim of
this prospective study was to compare a new “leadless” ambulatory monitor with a standard event monitor in the evaluation
of patients with palpitations.
Methods
Eighteen consecutive patients (11 female, 56 ± 16 years) referred for evaluation of palpitations were provided with both a
standard event monitor and a “leadless” monitor for 30 days. They were asked to record episodes of palpitations with both
monitoring devices.
Results
All 18 individuals were compliant with the “leadless” monitor for the 30-day period while only 14 (78%) patients were compliant
with the standard event monitor (p = 0.10). During a combined monitoring period of 563 days, 159 symptomatic episodes were recorded with the “leadless” ECG
monitor (8.8 ± 9.7 per patient, range 1–35) and 169 symptomatic episodes were recorded with the event monitor (12 ± 8.3 per
patient, range 1–33) (p = NS). The “leadless” ECG monitor recorded arrhythmias in 13 of 18 patients (72%) and the standard event monitor recorded
arrhythmias in 8 of 14 patients (57%) (p = NS).
Conclusion
The “leadless” ECG monitor is associated with high patient compliance and results in high quality ECG recordings. The diagnostic
yield of this monitoring system is equivalent to a standard event monitor.
Keywords Palpitations - ECG - Ambulatory Monitor - Arrhythmias
Dr. Daniel Scherr was supported by a research grant of the FWF Austrian Science Fund.