Modelling the Acquisition Geometry of a C-Arm Angiography System for 3D Reconstruction
| |
|
Modelling the Acquisition Geometry of a C-Arm Angiography System for 3D Reconstruction
Cristina Cañero3, 4 , Eduardo Nofrerías5, Josefina Mauri5 and Petia Radeva3, 4
| (3) |
Computer Vision Center, Edifici O Campus UAB, 08190 Cerdanyola, Spain |
| (4) |
Computer Science Department, Universitat Autònoma de Barcelona, Spain |
| (5) |
Hospital Universitari Germans Tries i Pujol, Badalona, Spain |
Abstract
This paper is concerned with the calibration of the intrinsic and extrinsic parameters of a X-ray angiographic system for
3D reconstruction of coronary vessel centerlines. We discuss whether the isocentric model is an accurate model for the movement
of the C-arm, and analyse the assumptions made by this model. We propose a methodology which takes into account these facts
and we propose for each model a calibration method. The calibration only needs a phantom that is commonly available by the
physicians, since it is used to calibrate geometrical distortion. Then, we test the models with the calibration grid, as well
as with a phantom imaged three months later. Experimental results show that by using the proposed models, it is not necessary
neither a calibration nor a refinement at each acquisition.