Transpalpebral measurement of intraocular pressure using the TGDc-01 tonometer versus standard Goldmann applanation tonometry

Annette Lösch, Alexander Scheuerle, Volker Rupp, Gerd Auffarth and Matthias Becker

View Related Documents

Abstract

Background  

A recently developed digital tonometer for transpalpebral intraocular pressure (IOP) measurement, distributed by Corneal, Inc., allows the noninvasive measurement of IOP for screening purposes.

Method  

We measured the IOP of 218 eyes in 109 patients of the Interdisciplinary Uveitis Center of the University of Heidelberg with intact corneal epithelium. IOPs were measured first with the TGDc-01 tonometer, and then by means of Goldmann tonometry. IOPs were recorded by two independent examiners. The mean of three measurements obtained with the TGDc-01 was taken, whereas only one measurement was performed with the Goldmann tonometer.

Results  

The mean difference between the TGDc-01 and Goldmann measurements was 3.7 mmHg. The standard deviation of the differences was ±4.06 mmHg. Thus measurements acquired with the TGDc-01 may range 4.4 mmHg above or 11.8 mmHg below the values given by Goldmann tonometry.

Conclusion  

The IOP values obtained with the TGDc-01 were in poor agreement with Goldmann tonometry. We found a higher variation as well as a bias towards lower IOP values with the TGDc-01. It is a question of clinical judgement as to how far these deviating measurements can be accepted for screening purposes. Because the IOPs obtained with the TGDc-01 are generally lower and less accurate than those obtained with the Goldmann tonometer we believe that the TGDc-01 is not a reliable tool for IOP measurement in clinical routine.
This work was presented at the 101st meeting of the German Society of Ophthalmology, Berlin, 2003