In the last few years there has been great interest in minimally invasive plate osteosynthesis (MIPO) in the treatment of
humeral shaft fractures. None of these studies showed the anatomical relationship between the radial nerve and the material
of the implant in vivo. We performed postoperative ultrasonographic measurement of the distance between the radial nerve and
the material implanted using the MIPO technique. Nineteen patients underwent postoperative ultrasound examinations. Group
A comprised midshaft fractures and group B distal third fractures. The point of greatest proximity between the radial nerve
and the implant was measured. In group A the distance was between 1.6 and 19.6 mm (mean: 9.3 mm) and in group B between 1.0
and 8.1 mm (mean: 4.0 mm). The ultrasound findings reveal that the radial nerve is quite close to the implant material, especially
in the transition between the third and fourth quarters of the humeral shaft.
Ce travail est l’étude par ultrasonographie des rapports entre le nerf radial et la plaque implantée par technique mini invasive
(MIPO) dans le traitement des fractures de la diaphyse humérale. Dix neuf patients sont étudiés avec fracture à mi-diaphyse
pour le groupe A et fracture du 1/3 distal pour le groupe B. Le point de plus grande proximité entre l’implant et le nerf
a été mesuré: entre 1,6 et 19,6 mm pour le groupe A (moyenne 9,3 mm) et entre 1,0 et 8,1 mm pour le groupe B (moyenne 4,0 mm).
L’étude montre la proximité entre le nerf radial et le matériel surtout à l’union 3/4 proximal et 1/4 distal.