A good deal has been reported about the outcomes of massive distal femoral endoprosthetic reconstructions after tumor resection. Recent studies have described distal femoral endoprosthetic survivorship at 10 years to range upward of 80% [6, 31, 39]. Less is known about proximal tibial oncologic reconstructive outcomes for a number of reasons. First, although the incidence of proximal tibial tumors ranks next to distal femoral lesions, the latter remain roughly twice as common [48]. Second, amputation is more often necessary for proximal tibial neoplasms because of neurovascular involvement. Third, the comparatively greater challenges posed by the reestablishment of extensor mechanism function after proximal tibial resection have led to the use of a variety of tibial reconstructive methods, including arthrodesis [12], allografts [10, 14, 43], and alloprostheses [5, 17, 51] with relatively fewer endoprosthetic reconstructions. Furthermore, although intermediate-term (ie, 5–10 years) prosthetic survivorship of a variety of distal femoral implants is reasonably predictable and acceptable [6, 33, 41, 45], tibial endoprosthetic reconstruction is more challenging principally because of the biomechanical demands placed on tibial stems. Two- to 5-year failure rates resulting from aseptic loosening have ranged from 3% to 46% in a variety of series [19, 22, 24, 26, 29, 34, 35].
Compressive osseointegration technology was developed in an attempt to provide secure, long-term anchorage of oncologic endoprostheses by using a spring-loaded device to achieve compliant prestress fixation, thus avoiding complications of stress shielding and particle-induced osteolysis [8, 15, 27, 36]. Initial distal femoral radiographic results have confirmed progressive bone hypertrophy at the prosthetic interface [3] and early clinical comparisons with cemented stems [4] have demonstrated equivalent prosthetic survivorship at 2 years.
For primary tumor proximal tibial resections managed with compressive osseointegration reconstructions, the purposes of this study were to determine (1) the rates of local control and prosthetic survival; (2) the frequency and nature of surgical complications; (3) and the outcome of prosthetic revision.
I retrospectively reviewed 16 patients with resected malignancies reconstructed with a proximal tibial Compress® device (Biomet, Inc, Warsaw, IN) between April 1998 and September 2006. There were seven males and nine females with an average age of 18 years (range, 12–42 years). Diagnoses included osteosarcoma (12), Ewing’s sarcoma (two), chondrosarcoma (one), and undifferentiated sarcoma (one). Distal femoral Compress® devices, initially introduced in 1993, were granted US Food and Drug Administration clearance in December 2003. On a custom off-label basis, Compress® implants have been available for proximal tibial reconstructions since 1998. The minimum followup was 2 years (mean, 4.5 years; range, 2–10.3 years). No patient was lost to followup. Prior Institutional Review Board approval was obtained for this retrospective review.
Previously published methods of proximal tibial resection and endoprosthetic reconstruction were followed [7, 13, 16, 18, 23, 28, 30, 36, 50]. A sufficient amount of proximal tibia was removed to achieve negative surgical margins. Resection length averaged 17 cm (range, 13–24 cm); remaining distal tibial segments averaged 20 cm (range, 11–25 cm); percentage of tibia resected averaged 46% (range, 35%–65%). The tibial canal was reamed to allow placement of a 10-mm anchor plug and a centering sleeve of at least 12 mm diameter. Compression force (in pounds) was as follows: 400 (five), 600 (10), and 800 (one); two short custom spindles were used. Over time, we have empirically preferred 600-pound small, short hydroxyapatite spindles for tibial reconstructions. Twelve patients underwent extensor mechanism reinforcement and soft tissue coverage with a gastrocnemius flap.
Continuous passive motion of the knee was begun 48 hours after surgery or as soon as permitted by the plastic surgery staff. Quadriceps sets and straight leg raising exercises were begun 2 weeks after surgery, but progressive resistive exercises were not undertaken. In contrast to rehabilitation of patients with cemented stems, for which full weight can be borne immediately, weightbearing on the tibial Compress® implant was withheld for 3 months postoperatively, after which time weightbearing was advanced at a rate of 25% of body weight per week.
Followup visits for routine clinical and radiographic examinations generally occurred at 2, 6, and 12 weeks and at 3-month intervals thereafter. Medical records were reviewed to obtain demographic data, including age, gender, diagnosis, and treatment information. Operative reports were studied to record technical factors, including implant length relative to remaining distal tibial length, compression force, and spindle length. Major complications such as aseptic loosening, infection, local recurrence, periprosthetic fracture, need for further surgery, metastatic disease, and death were recorded. Device-related mechanical failure was defined as the need for revision secondary to aseptic loosening.
I assessed postoperative radiographs at 3- to 6-month intervals for indications of technical error (pin malposition or migration) and for evidence of device-related failure (lucency at the bone-prosthetic interface, endosteal erosion, loss of compression distance, gross loosening, and implant breakage).
There were no local recurrences. Four patients had metastatic disease at last followup. Two deaths occurred; one patient died secondary to osteosarcoma, and one developed myelofibrosis.
Surgical complications included one early deep infection that ultimately resulted in prosthetic loosening and the need for an above-knee amputation. There were two late deep prosthetic infections; prosthetic retention was achieved with débridement and antibiotics. There were two nondisplaced tibial fractures, not associated with the hardware, which were treated nonoperatively. There were no prosthetic fractures or other forms of mechanical breakage. All patients were able to walk without an assistive device.
The rationale of this study was to better understand the short-term outcomes of compressive osseointegration technology when used for reconstruction of massive defects after primary oncologic resection. The primary aims were the determination of prosthetic survival and management of revisions. The secondary aims were documentation of local control and surgical complications.
Limitations of this study include the small number of patients, the lack of control subjects, and the limited followup. The limited study size makes statistical analysis difficult, especially in terms of meaningful calculation of implant survival analysis. With average 4.5-year followup, the data in this study should be viewed as preliminary with results needing to be further developed to substantiate conclusions regarding prosthetic longevity. However, given the relative infrequency of primary oncologic proximal tibial reconstructions, this review of 16 patients is one of the larger series of uncemented devices and the only one that relates to compressive osseointegration technology. Except for the two patients who died, all of the patients reported continue to be examined regularly so the longer-term durability of the implant can be determined.
Effective reconstruction of massive defects after resection of proximal tibial neoplasms is challenging for several reasons. First, as compared with femoral presentations, the size of tibial tumors and the close proximity of surrounding neurovascular structures often render decision-making regarding limb salvage difficult. As compared with distal femoral tumors, achieving local control of tibial lesions more often necessitates amputation. The correspondingly fewer patients who do receive prosthetic reconstructions may still be at high risk for local recurrence. Although not directly a product of osseointegration technology, the finding that all patients in this study demonstrated local control is nonetheless reassuring in terms of validating a limb salvage approach in a carefully selected population.
|
Study |
Year |
Device description |
Patients (number) |
Average followup (years) |
Mechanical failure rate |
Comment |
|---|---|---|---|---|---|---|
|
Jeys et al. [25] |
2008 |
Cemented custom constrained and rotating hinge |
136 |
10 |
31.4 |
Mechanical failure rate (31.4%) |
|
Myers et al. [34] |
2007 |
Cemented custom constrained hinge |
95 |
5 |
16 |
Aseptic loosening (16%) plus five prosthetic fractures |
|
Myers et al. [34] |
2007 |
Cemented custom rotating hinge, hydroxyapatite collar |
99 |
5 |
3 |
Aseptic loosening (3%) plus seven prosthetic fractures |
|
Flint et al. [19] |
2006 |
Uncemented modular constrained hinge |
44 |
3 |
6.8 |
Stem breakage (two); rotational instability (one) |
|
Gosheger et al. [20] |
2006 |
Uncemented modular rotating hinge |
42 |
4 |
9.5 |
Aseptic loosening (three); stem breakage (one) |
|
Ahlmann et al. [2] |
2006 |
Cemented modular rotating hinge |
30 |
3 |
10 |
Fatigue fracture (two); aseptic loosening (one) |
|
Torbert et al. [44] |
2005 |
Cemented modular rotating hinge |
26 |
5 |
19.2 |
Mechanical failure (three); aseptic loosening (one); dislocation (one) |
|
Natarajan et al. [35] |
2003 |
Cemented custom constrained and rotating hinge |
133 |
5 |
10.5 |
Fracture revision (seven); aseptic loosening (five); disassembly (one); bending (one) |
|
Kawai et al. [26] |
1999 |
Uncemented (five) and cemented (two) rotating hinge |
7 |
2 to 7 |
28.6 |
Tibial yoke breakage (two) |
|
Horowitz et al. [24] |
1991 |
Cemented custom constrained hinge |
16 |
5 |
25 |
Aseptic loosening (three); articulation fracture (one) |
A third problematic aspect of tibial reconstruction involves soft tissue coverage and the consequent risk of prosthetic infection. One patient without flap coverage in this series developed an infection that ultimately necessitated an amputation; the two other patients who developed a deep infection retained their implants, perhaps in part because they had received a flap as part of their index procedure. The utility of the gastrocnemius flap in decreasing the risk of tibial prosthetic infection and in improving extensor mechanism reconstitution serves to makes its use routine [7, 11].
A final challenge of tibial reconstruction is the frequent need to salvage short metadiaphyseal fragments remaining after tumor resection or revision. The “bone-sparing” nature of the Compress® device (as little as 43 mm of bone can be implanted) and the relative ease of revision after infection, fracture, or mechanical loosening (the device is readily removed and as little as 1 cm of additional bone needs to be resected at the time of reimplantation) are highlighted by the successful revision of the sole case of mechanical failure (Fig. 1) [36].