After TKA, some patients present with extensive wear and osteolysis around well-fixed and well-aligned components. A similar scenario of osteolysis and polyethylene wear occurs around well-fixed modular acetabular components. In the hip, the technique of managing this scenario with liner exchange and bone grafting of the osteolytic lesions has been reported [7, 8]. One study of patients (24 hips) treated in this manner at mean 36 month followup reported 17 of 18 lesions regressed or resolved, and only two patients had additional revisions [9]. A longer term study of liner exchange for wear and osteolysis of the hip reported a decrease in the rate of lysis as well as hips remaining radiographically stable at 6.2 year followup [11].
Isolated liner exchange after TKA has had limited success when the indications have been for all causes. One series reported a 25% rerevision rate at mean three year follow-up for tibial insert exchanges for wear or instability [1]. Another series reported 17% failure of isolated insert exchanges for wear and other diagnoses at average 54 months [3]. A more recent study of isolated liner exchange and grafting for wear and osteolysis following TKA reported a 16.2% failure rate after a mean 44 month followup [5]. For the above reasons, this approach has not gained popularity in cases of extensive osteolysis and polyethylene wear after TKA.
With the use of modular tibial components, cases of extensive osteolysis around fixed total knee components have increased, presumably related to backside polyethylene wear as well as poor quality (ie, gamma irradiated in air) polyethylene. Occasionally, the lesions are so extensive that major structural allografts or even tumor-type prostheses are needed to reconstruct the defects. In cases in which the components were secure and well aligned, the authors decided to débride the lesions and fill the defects with bone grafting or bone graft substitute.
The purpose of this study was to evaluate the technique of liner exchange and bone grafting for cases of wear and osteolysis around TKAs in which the components were well-fixed and well-aligned to determine (1) rerevision rates; (2) fate of the bone graft; (3) radiographic loosening rates; and (4) functional scores.
We retrospectively reviewed 22 selected patients who underwent 25 total knee revisions with only exchange of the tibial liner and bone grafting of osteolytic lesions at four institutions (University of Iowa, Rush University, Des Moines Orthopaedic Surgeons, and Washington University) over a 10.5-year period (October 1995 through May 2006). The indication for revision TKA surgery was wear and osteolysis. All knees were well-fixed as determined radiographically with the final determination of fixation made intraoperatively, and all were well-aligned (mechanical axis of neutral ± 1 to 2° and anatomic alignment of 5–9° of valgus). We excluded any case having revision of either the femoral or tibial component (regardless of reason, including problems with the capturing mechanism of the tibial tray). In all of these patients, the alternate form of surgery would have been with structural allografts or extensive metal augmentation in the revision reconstruction. The typical scenario for these patients was that they returned for followup after not having been seen for several years, or they were new patients who had not been followed by their original surgeon. They presented with large osteolytic lesions and were not lesions that had been closely followed over time to decide when treatment was warranted. Twenty-five TKAs (less than 1% of the knee revisions performed by the seven surgeons) met the criteria and were included in the study. There were 17 females and five males. The average age at the time of the index liner exchange was 63 years (range, 45–85 years), which was on average 9 years (range, 3–14 years) after the primary TKA. The average BMI for the patients was 33 kg/m2 (range, 23–48 kg/m2). The minimum clinical followup was 22 months postliner exchange (average, 61 months; range, 22–142 months). The minimum radiographic followup was 22 months postliner exchange (average, 59 months; range, 22–130 months) with one knee excluded as a result of radiographic followup of less than 1 year. No patients were lost to followup. No patients were recalled specifically for this study; all data were acquired from medical records and radiographs.
|
Implant design |
Number (n = 25) |
Percent |
|---|---|---|
|
AMK* |
4 |
16 |
|
Insall Burstein† |
2 |
8 |
|
Miller Galante† |
2 |
8 |
|
PCA‡ |
2 |
8 |
|
PFC* |
12 |
48 |
|
Scorpio§ |
2 |
8 |
|
Whitesides Cementless|| |
1 |
4 |
|
AP projection |
Osteolysis |
|---|---|
|
Femur |
9 knees |
|
21.3 cm2 average (range, 5–48 cm2) |
|
|
Location |
|
|
Medial 5 |
|
|
Lateral 2 |
|
|
Global 2 |
|
|
Tibia |
19 knees |
|
10.2 cm2 average (range, 1–37.2 cm2) |
|
|
Location |
|
|
Medial 11 |
|
|
Lateral 2 |
|
|
Global 6 |
|
Lateral projection |
Osteolysis |
|---|---|
|
Femur |
12 knees |
|
21.7 cm2 average (range, 9–54 cm2) |
|
|
Location |
|
|
Anterior 0 |
|
|
Posterior 2 |
|
|
Global 10 |
|
|
Tibia |
12 knees |
|
9.3 cm2 average (range, 1–31.5 cm2) |
|
|
Location |
|
|
Anterior 8 |
|
|
Posterior 3 |
|
|
Global 1 |
|
Graft source |
Number |
Percent |
|---|---|---|
|
Allograft |
11 |
44 |
|
Allograft and commercial graft |
8 |
32 |
|
Commercial graft |
4 |
16 |
|
Allograft and methylmethacrylate |
2 |
8 |
Followup included clinic visits at 6 weeks, 3 months, and 1 year postoperatively, with AP, lateral and Merchant radiographs taken at the 3 month followup. Further followup is at an interval of every 2–3 years, at which time radiographs are repeated. If patients are having problems or symptoms, they are followed more frequently. Knees were evaluated clinically for need for rerevision/reoperation (ie loosening, lack of graft incorporation, or component subsidence) and with the Knee Society score [6] at final followup.
Three of us (JJC, ERR, SSL) evaluated all AP and lateral radiographs with final determination by consensus. Radiographic analysis consisted of evaluation for loosening [4], osteolysis [10], and bone graft incorporation. Bone graft incorporation was determined by noting differences in defect filling on comparison of postoperative and final followup radiographs. Full incorporation was defined as no difference in defect filling, mostly incorporated was defined as a small amount of difference in graft resorption (ie < 1 cm2), and partial incorporation was defined as at least 25% resorption of graft.
At mean 61-month followup, one patient (one knee) underwent reoperation for aseptic loosening of the femoral and tibial components with osteolysis of both the femur and tibia as well as wearthrough of the metal-backed patellar component. The patient originally had the index liner exchange and bone grafting for wear and tibial osteolysis. The bone graft showed little to no incorporation 8 years postliner exchange. One additional patient (one knee) had arthroscopic débridement approximately 2 years postliner exchange for peripatellar fibrosis. No other patients had further reoperations.
|
Lesions |
Number |
Percent |
|---|---|---|
|
Femoral lesions (13) |
||
|
Fully incorporated |
9 |
69.2% |
|
Mostly incorporated |
2 |
15.4% |
|
Partially incorporated |
2 |
15.4% |
|
No incorporation |
0 |
|
|
Tibial lesions (20) |
||
|
Fully incorporated |
12 |
60% |
|
Mostly incorporated |
2 |
10% |
|
Partially incorporated |
5 |
25% |
|
No incorporation |
1 |
5% |
At a mean of 59 months, no additional components demonstrated evidence of radiographic loosening (component position change or complete or near complete circumferential radiolucent lines) other than the one knee that was revised. All other components were well fixed at final followup with no evidence of loosening.
At last followup, the mean clinical and functional Knee Society scores were 90 (range, 58–100) and 73 (range, 40–100) points, respectively.
After TKA, some patients present with extensive wear and osteolysis around well-fixed and well-aligned components. A similar scenario of osteolysis and polyethylene wear occurs around well-fixed modular acetabular components. Several authors have reported liner exchange and bone grafting of periacetabular osteolytic lesions [7, 8], but not liner exchange and substantial osteolysis around a TKA. We therefore evaluated the technique of liner exchange and bone grafting for cases of wear and extensive osteolysis around TKAs in which the components were well-fixed and well-aligned to determine (1) rerevision rates; (2) fate of the bone graft; (3) radiographic loosening rates; and (4) functional scores.
There are several limitations to the present study. First, we had a relatively small number of cases. This is not a widely used technique and, when combined with the specific indication and exclusion criteria, resulted in a very few cases. Second, multiple surgeons at multiple institutions contributed to the study to obtain adequate numbers. However, the larger number of surgeons likely means the findings are more generalizable. Further, although this is a multicenter study, the indications and exclusion criteria for this procedure were similar for all surgeons. Third, different types of prostheses were used in the total knees in this series. Because this study cohort involved revisions of the liner and bone grafting only, we were constrained by the fact that the prostheses used in the initial situation were dictated at the time of primary TKA. Fourth, there were a number of different types of bone grafting material used. The decision regarding what type of material to use was left to the discretion of the surgeon performing the procedure. Ideally, it would have been better if all surgeons used the same material, but this could not be the case because not all institutions have the same materials available. Fifth, we are aware of the possible limitations of examining osteolytic lesions on radiographs. Radiographs reportedly underestimate the extent of osteolysis [2]. While it would have been better to do further imaging, such as CT evaluation, it was not available in this retrospective study. Sixth, alignment was determined on an AP radiograph since long cassettes were not always available. However, alignment was reconfirmed intraoperatively. Seventh, quantification of articular and backside wear was not performed on the removed tibial liners. Finally, the results reported are the initial results of our combined experience at average 5-year followup. Further followup is warranted.
At a mean of 61 months followup, only one case (4%) was revised for loosening 8 years after the initial revision. This compares more favorably than the other rerevision rates reported in the literature for liner exchange revisions in TKA, with rates reported as high as 25% at mean 3 year followup [1, 3, 5]. One key difference is that those previous studies had indications for liner exchanges beyond just wear and osteolysis and included instability and stiffness among other diagnoses. To our knowledge, there are no reported series of liner exchange only and bone grafting for cases of extensive osteolysis around TKAs.
Fully or mostly incorporated graft occurred in 11 of 13 (84.6%) femoral lesions and 14 of 20 (70%) tibial lesions, respectively. This is similar to other published results for liner exchanges in TKAs with 97% of osteolytic lesions demonstrating no progression at mean 44 month followup [5] and THAs with 94% of osteolytic lesions regressing or resolving at 3 year followup [9].
Mean clinical and functional Knee Society scores were 90 (range, 58–100) and 73 (range, 40–100) points, respectively, which was comparable to the mean 92 reported at mean 44 month followup in another series [5].
The present series of 25 cases with extensive osteolysis around well-fixed and well-aligned components in which débridement of the osteolytic lesions, bone grafting of the lesion, and liner exchange with appropriate ligament balancing has demonstrated excellent results at an average of 61 months with only one case (4%) rerevised 8 years after the initial revision and no additional component loosening. Most of these cases would have required massive allografts or metal augmentation at revision surgery. The authors continue to selectively use this approach and recommend that it be included in the options at the time of revision TKA. The only caveats are that the strict indication and exclusion criteria should be adhered to and that this is not a common scenario (< 1% of the revisions performed by the surgeons involved).