The aim of this study was to evaluate the validity and reproducibility of breath-hold and respiratory triggered phase-contrast
(PC) MR imaging techniques in the measurement of renal artery blood flow. In 12 healthy subjects cardiac-gated PC flow measurements
were obtained in the renal arteries using a breath-hold and a respiratory-triggered technique. The flow measurements were
repeated in each renal artery separately. Comparison between the sum of flow measurements in the renal arteries and the difference
in aortic flow measurements above and below the renal arteries served as an internal control. The flow measurements showed
a good reproducibility both with the breath-hold (
r = 0.92,
p < 0.0001) and with the respiratory-triggered (
r = 0.91,
p < 0.0001) technique. The validity of both methods was good and there was no statistically significant difference. Reproducible
quantitative measurements of renal artery blood flow are possible with respiratory controlled, cardiac-gated, PC MR imaging.
Key words: Renal arteries – MR imaging – Flow dynamics – Phase imaging