Radiofrequency ablation in the liver close to the bile ducts: can intraductal cooling offer protection?

U. Jersenius, D. Arvidsson, J. Lindholm, S. Anttila and A. Elvin

View Related Documents

Abstract

Background  

One complication of radiofrequency ablation (RFA) of the liver is biliary duct damage. Intraductal cooling (IDC) has been proposed as a means of protection.

Methods  

In the first experiment, designed to evaluate the influence of IDC on the RFA procedure per se and on lesion formation, lesions were created in vivo in pig liver with and without IDC. The RFA needle was placed with a 1.5-cm safety margin from the bile ducts. In the second experiment, designed to evaluate the potential protective effects of IDC, lesions were created close to a bile duct with and without IDC.

Results  

With the safety margin, the RFA parameters and lesion size were not negatively affected by IDC. Microscopic examination revealed that IDC had a protective effect in most of the lesions created close to a bile duct.

Conclusions  

The IDC procedure was feasible and had no negative effect on the RFA procedure or the lesions. However, the protective effect of IDC was not statistically significant (p = 0.12).

Keywords  Radiofrequency ablation - Bile duct damage - Intraductal cooling - Liver