Object
Internal MRI coils have important applications in diagnostic and interventional studies. Since they can be placed very close
to the region of interest in the body, they are favored over external coils in applications where high-resolution images are
required. In this paper it is demonstrated that ultimate intrinsic SNR (UISNR) and the optimum coil sensitivity solutions
can be used to make new coil designs with higher intrinsic SNR.
Materials and methods
In this study, UISNR, which is the maximum attainable value of the intrinsic SNR, is used as a measure of performance and
as a design criterion. As an example, a novel endorectal MRI coil is designed. The design is tested with phantom and patient
studies.
Results
An endorectal coil is built to demonstrate the effectiveness of the design strategy. ISNR of the endorectal coil approximates
the UISNR to 72%.
Conclusion
An internal coil design method that takes advantage of the UISNR and optimum coil sensitivity calculations was presented.
This method can also be used to design better internal MRI coils for different applications.
Keywords Internal MRI coils - Intrinsic SNR - Ultimate intrinsic SNR