Objective. The accuracy and precision of the new IRMA® (Immediate Response Mobile Analysis System, Diametrics, Inc.®, St. Paul, MN)
handheld blood gas analyzer was compared with that of two benchtop blood gas analyzers. The IRMA consists of a notebook-sized
machine and disposable cartridges, each containing a pH, a CO
2 and an O
2 electrode, and provides bedside (point-of-care) blood gas analysis.
Methods. A total of 172 samples (arterial and mined venous) were obtained from 25 informed, consenting patients undergoing cardiopulmonary
bypass. The pH, PCO
2 and PO
2 of each sample was determined on four blood gas analyzers: NOVA Statlabs Profile 5 (NOVA Biomedical, Waltham, MA), the ABL-50
(Radiometer, West Lake, OH), and two IRMA machines. Linear regression and bias ± precision were determined, comparing each
of the analyzers with the NOVA.
Results. All three machines showed a similar, high degree of correlation with the NOVA for pH, PCO
2, and PO
2. The bias and precision of the IRMA machines compared with the NOVA was similar to that of the ABL compared with the NOVA
for pH (NOVA:ABL −0.005 ± 0.011; NOVA: IRMA 1 = 0.0026 ± 0.025; NOVA: IRMA 2 = 0.0021 ± 0.025), for PCO
2 (NOVA:ABL = −1.4 ± 1.3 mmHg; NOVA: IRMA 1 = −1.3 ± 1.9 mmHg; NOVA: IRMA 2 = −1.2 ± 2.1 mmHg) and PO
2 (NOVA:ABL = 3.6 ± 21.1 mmHg; NOVA: IRMA 1 = 3.4 = 19.9 mmHg; NOVA: IRMA 2 = 6.3 ± 20.9 mmHg). The bias found for pH, PCO
2, and PO
2 was not affected by extremes of temperature (range 25.5–40°C) or hematocrit (range 11–44%) for any machine.
Conclusions. The new technology incorporated in the IRMA blood gas analyzer provides results with an accuracy that is similar to that
of benchtop analyzers, but with all of the advantages of point-of-care analysis.
Key words Blood gas analyses - Monitoring: oxygen, carbondioxide, pH - Equipment: point of care blood gas analyzer